Corsano Vital Parameters

Clinically validated and certified analytics

The software in the CardioWatch 287-2 System generates Vital Parameters with Corsano proprietary algorithms established within the Corsano Bracelet and Cloud as well managing unadjusted measurements with external devices.

Vital Parameter

Algorithm

Validation

Certified CE-MDR

Pulse Rate (PR)

Corsano Proprietary in FW

CCN, CMC at rest and under motion

Saturation (Sp02)

Corsano Proprietary in FW

CMC at rest and under motion

Respiration Rate (RR)

Corsano Proprietary in FW

CMC at rest and under motion

Cuffless Non-Invasive Blood Pressure (NIBP)

Corsano Proprietary in
Cloud

Reinier de Graaf Hospital

Electrocardiography (ECG)

Corsano SQ + CardioLyse

Kiwa Dare + RECAMO

Skin Temperature (sTemp)

Surface probe in direct mode

VDE Prüf- und Zertifizierungsinstitut GmbH

Core Body Temperature (CBT)

Heatflux + GreenTeg Fever

Basel University Hospital

Steps (STEPS)

Corsano Proprietary in FW

Corsano, RadboudMC

Axillary Temperature (aTemp)

External Vivalink Axillary Temperature Sensor

FDA Cleared (FDA K162137)

Non-Invasive Blood Pressure (NIBP)

External Non-Invasive Blood Pressure Monitor Transtek TMB-2084-A

FDA Cleared (K220676)

Spirometry (SPIRO)

External MIR Spirobank

FDA Cleared (K072979)

Weight (WEIGHT)

External Weight Scale Transtek GBS-2012-B

FDA Listed (D1545656)

Corsano Algorithm summary

The Firmware running on the Corsano Bracelet measures raw data of the PPG, Accelerometer and Temperature sensors and runs algorithms to compute the vital parameters:

  • Pulse Rate
  • Oxygen Saturation
  • Respiration Rate
  • NIBP (requires Corsano Cloud)
  • Temperature
  • Activity
Described below the principles of operation per vital parameter:

Pulse Rate principal of operation

The Corsano Bracelet utilizes a Photoplethysmography (PPG) sensor that consists of Light Emitting Diodes (LEDs) and photodiodes to capture reflected light. PPG is a commonly used in determining Pulse Rate (BPM), Oxygen Saturation (SpO2) and Respiration Rate (RR).
The working principle of the PPG sensor is based on the emission of LED light which penetrates the skin and blood vessels. This light is then reflected and captured by the photodiodes in Corsano Bracelet to measure the blood stream.
The results of the PPG signal depend primarily on the flow of blood to the capillary vessels in each heartbeat. The waveform of the PPG signal indicates the changes in pulsatile blood flow from which the detection of signal peaks allows the calculation of peak-to-peak intervals. By determining the peak-to-peak distances between two subsequent PPG pulses, the algorithm derives the Pulse Rate of the patient.
The accelerometer (ACC) is utilized to compensate for movement artifacts. PPG and ACC are measured at 32Hz and data is processed by the algorithm to give a new reading of Pulse Rate every 28 seconds utilizing averaging to smooth pulse rate data and prevent inappropriate and transient artifacts from affecting stability of results.

Parameter

Specification

Sensing Method

PPG, ACC

Measurement Range

30 bpm to 240 bpm

Resolution

1 bpm

Accuracy

1.95 Arms

Bias

0.44 (0.42 - 0.46) bpm

Upper LoA

4.17 (4.15 - 4.19) bpm

Lower LoA

-3.3 (-3.32 - -3.28 bpm

Data Update Period

28 seconds

Pulse Rate Specification

The accuracy of the pulse rate algorithm was measured by Complementair Medisch Centrum (CMC) Europe, Belgium, in a controlled study with 24 healthy adult volunteers in accordance with ISO 80601-2-61:2017 at rest and under motion.

Spo2 principal of operation

Corsano Bracelet utilizes the Photoplethysmography (PPG) sensor photodiodes to capture reflected light calculates the patient’s oxygen saturation (SpO2) using pulse oximetry, a non-invasive technique to monitor oxygenation. It monitors the percentage of hemoglobin that is oxygen-saturated. The working principle is spectrophotometry: the relative absorption of red (absorbed by deoxygenated blood) and infrared (absorbed by oxygenated blood) light of the systolic component of the absorption waveform correlates to arterial blood oxygen saturations. Two light-emitting diodes, red with wavelength of 660 nm and infrared with a wavelength of 880 nm, are positioned so that they are opposite their respective photodiodes through 5-10 mm of tissue. Absorption of light at these wavelengths differs significantly between blood loaded with oxygen and blood lacking oxygen. Oxygenated hemoglobin absorbs more infrared light and allows more red light to pass through. Deoxygenated hemoglobin allows more infrared light to pass through and absorbs more red light. The accelerometer (ACC) is utilized to compensate for movement artifacts. Measurements of relative light absorption and ACC are measured at 32Hz and data is processed by the algorithm to give a new reading of SpO2 every 28 seconds utilizing averaging to smooth SpO2 data and prevent inappropriate and transient artifacts from affecting stability of results.

Parameter

Specification

Sensing Method

Pulse Oximetry

Measurement Range

0% to 100%

Resolution

1%

Accuracy

70-100%: 1.39 Arms


90-100%: 1.34 Arms
80-90%: 1.42 Arms
70-80%: 1.45 Arms

Bias

70-100%: 0.23 (0.07 – 0.39)
 
90-100%: 0.01 (-0.23 – 0.25)
80-90%: 0.26 (-0.02 – 0.55)
70-80%:
0.52 (0.22 – 0.82)

Upper LoA

70-100%:  2.93 (2.77 – 3.08)
 
90-100%: 2.64 (2.4 – 2.88)
80-90%: 2.99 (2.71 – 3.28)
70-80%:
3.16 (2.87 – 3.46)

Lower LoA

70-100%: -2.47 (-2.62 - -2.38)
 
90-100%: -2.62 (-2.86 - -2.38)
80-90%: -2.47 (-2.75 - -2.18)
70-80%:
-2.12 (-2.42 - -1.83)

Data Update Period

28 seconds

Nominal Wavelength

Red: 660nm
Infra-Red: 880nm
Green: 525nm

Spo2 Specification

The accuracy of the pulse oximetry function was measured by Complementair Medisch Centrum (CMC) Europe, Belgium, in a controlled induced hypoxia study using 24 healthy adult volunteers in accordance with ISO 80601-2-61.

Bland-Altman plot for CardioWatch 287-2 SpO2 and CO-oximetry SaO2 in the invasive study (correlation was 98.8%, N=300).

The following shows correlation and Bland-Altman between CardioWatch 287-2 and the Nellcor PM10 from the non-invasive study (N=94572)

Correlation and Bland-Altman SpO2 vs Nellcor PM10

Respiration rate principal of operation

The Respiration Rate is the number of breaths a person takes per minute (RR). Corsano Bracelet utilizes a proprietary time-frequency algorithm to extract the respiratory-induced intensity, amplitude and frequency variation signals from the photoplethysmography (PPG) signal.

The accelerometer (ACC) is utilized to compensate for movement artifacts. PPG and ACC are measured at 32Hz and data is processed by the algorithm to give a new reading of Respiration Rate every 28 seconds utilizing averaging to smooth pulse rate data and prevent inappropriate and transient artifacts from affecting stability of results.

Parameter

Specification

Sensing Method

PPG, ACC

Measurement Range

6 brpm – 60 brpm

Resolution

1 brpm

Accuracy

Laying in supine position: <1 Arms
Laying in prone position: <2 Arms
Laying on the side: <1 Arms
Sitting 90° position: <2 Arms
Sitting 45° position : <1 Arms
Standing upright: <2 Arms
Shallow and fast breathing:
<2
Arms

Bias

Overall: 0.03 (0.02 – 0.04) brpm

Upper LoA

Overall: 1.82 (1.81 – 1.83) brpm

Lower LoA

Overall: -1.76 (-1.77 – -1.75) brpm

Data Update Period

28 seconds

Apnea Detection

No

Respiration Rate Specification

The accuracy of Corsano Bracelet Respiration Rate measurement was determined by Complementair Medisch Centrum (CMC) Europe, Belgium.

Cuffless NIBP

Advanced photoplethysmography (PPG) signal analysis extracts the characteristics of the reflection waveforms. They are then fed to an artificial intelligence (AI) model. Thus, the model can predict systolic and diastolic blood pressure values.

The Corsano Bracelet makes use of the PPG sensor, which includes the emission of light by LEDs to the skin and the measurement of reflected light by photodiodes. The PPG signal is strongly dependent on the amount of light that is absorbed by the arterial blood in the tissue and reflectance on arterial/arteriole blood flow. As such, the PPG signal contains a pulsatile component that fluctuates with the blood volume changes in the peripheral tissue introduced during each cardiac cycle. This waveform contains information that is either directly or indirectly related to blood pressure. By extracting this information from the waveform and comparing it to previously acquired data from patients’ datasets, a prediction is made of blood pressure. The Corsano Bracelet extracts information from 6 simultaneously acquired waveforms, combining 3 wavelengths, 6 LEDs and 2 photodiodes.

From the PPG segments, features are extracted, which are used in the Machine learning model. To get a valid prediction, a combination of statistical features, time and frequency domain features, demographic features, first/second derivative features, width related PPG features and features from the PPG signal are used.

The Corsano CardioWatch 287-2 System was validated against the harmonized standard ISO 81060-2:2019 and AAMI/ESH/ISO collaboration statement "A Universal Standard for the Validation of Blood Pressure Measuring Devices" 2018. for Non-Invasive Blood Pressure measurement devices, in a clinical investigation on 97 subjects.

Correlation between Corsano Non-Invasive Blood Pressure measurement and the reference (Invasive measurement, Fysicon)
Bland-Altman plot comparing the Corsano Non-Invasive Blood Pressure measurement and the reference (Invasive measurement, Fysicon)

Parameter

Specification

Sensing Method

Derived from PPG signal

Measurement Range

60-230 mmHg (Systolic)
40-130 mmHg (Diastolic)

Resolution

1 mmHg

Accuracy

<5 mmHg

Bias

± 2 mmHg

Data Update Period

28 seconds

Non-Invasive Blood Pressure Specification

Corsano NIBP 24hr

Correlation Predicted DIA & SYS vs ABPM

Corsano NIBP under HIIT Exercise

Correlation Predicted DIA & SYS during High Intensity Interval Training (HIIT)

ECG

The Corsano CardioWatch 287-2 Bracelet implements a 1-lead electrocardiographic (ECG) system featuring electromagnetic interference (EMI) filtering, internal lead biasing, right-leg drive and lead-off detection. 

The system is composed of 3 electrodes. Two electrodes are located on the case back of the Bracelet, contacting the wrist skin through dry contacts. They are the positive ECG electrode and the right-leg electrode. The right-leg electrode enables to increase the rejection of noise and common-mode signals. The third electrode is the metal frame on the top of the Bracelet. This is the negative ECG electrode.

The patient must wear the Bracelet on one wrist and touch the top frame with the other hand to enable the ECG system to measure the electrical heart signal flowing from one arm to the other.

ECG Electrodes and signal

The electrodes are internally connected to the low noise and high-performance ECG circuit comprising the latest generation of analog front-end (AFE) integrated circuits. The electronics layout implements noise reduction and immunity techniques to achieve the performances required by medical standards. The ECG circuit measures the heart electrical signal and rejects noise and common-mode signals, thanks to analogue and digital filters and right-leg drive. Lead-off detection enables to verify the proper contacts of electrodes and improves the usability of the Bracelet.

The Corsano CardioWatch 287-2 ECG system was validated according to the harmonized standard EN-IEC 60601-2-47:2015 for performance and safety of ambulatory ECG systems. The tests were performed by the accredited laboratory Kiwa Dare in Woerden, NL.

Parameter

Specification

Sensing Method

1-lead ECG with right-leg noise rejection

Electrodes

Contact Dry Electrodes

Sampling Frequency

256 Hz

Signal range

±1300 mV

Frequency range

0.05 – 55 Hz

Resolution

15.3 ENOB (effective Number of Bits)

Common-Mode Rejection Ratio

110 dB (at 50 Hz and 60 Hz)

Input Impedance

1 GΩ

Performance

EN-IEC 60601-2-47

ECG Specification

Skin temperature principal of operation

Corsano Bracelet monitors skin temperature in direct mode from a sensor in the wearable device. Skin temperature is presented in °C, or in °F as configured in Settings/Profile.

Parameter

Specification

Sensing Method

Surface probe in direct mode

Measurement Range

34.0 °C to 42.0° C (93.2°F to 107.6°F)

Resolution

0.1°C

Accuracy

+/- 0.3°C (0.5°F)

Performance

Conform to ISO 80601-2-56

Data Update Period

28 seconds

Skin Temperature Specification

The accuracy of Corsano Bracelet temperature measurement was determined by the testing laboratory VDE Prüf- und Zertifizierungsinstitut GmbH, Offenbach, Germany in accordance with ISO 80601-2-56 - Medical electrical equipment Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement.

The Corsano Bracelet temperature measurement was compared to the reference Thermometer, Fluke Type 1524 with Probe 4298.

Core Body temperature principal of operation

CBT is one of the most relevant vital parameters to assess the state of the human body. Many processes in the human body are closely linked to CBT. However, CBT is also one of the most difficult parameters to monitor continuously and non-invasively.

A heat flux sensor in combination with PPG allows for reliable continuous monitoring of CBT as reported in prior scientific articles.

CardioWatch 287-2 integrates a heatflux sensor from GreenTeg and we combine its data with PPG derived BPM to improve accuracy.

CBT was validated versus electronic pills as a reference to assess the accuracy of the device and the algorithm. CardioWatch 287-2 raw measurement values were recorded during a clinical study with volunteers that were vaccinated for Covid, about half got a fever. Their body temperature was subsequently recorded and plotted against the temperature reference. The sensor signals were algorithmically combined to yield a single temperature output per day while at the same time compensating for physiological thermoregulation. The CBT algorithm is self-calibrating, meaning, after a certain calibration time, the algorithm will automatically determine several system parameters. From this moment on CBT can be calculated using these parameters. Following Bland-Altmann-plots represent the standard deviation from values to its reference.

Bland Altman CBT vs BodyCap ingestible CE Certified connected pill
Correlation CBT vs BodyCap ingestible CE Certified connected pill

Parameter

Specification

Sensing Method

Heatflux

Measurement Range

34.0 °C to 42.0° C (93.2°F to 107.6°F)

Resolution

0.1°C

Bias

0.14°C (0.18°F)

Stan dev.

0.34°C (0.61°F)

Lower LOA

-0.53°C (-0.95°F)

Upper LOA

0.82°C (1.47°F)

95% LOA

0.68°C (1.22°F)

Data Update Period

28 seconds

Core Body Temperature Specification

Steps

Steps are calculated by the proprietary algorithm based on the intensity and frequency of the patient’s acceleration data.

Accuracy of Steps with Corsano Bracelet were studied by RadboudMC versus Reference Device FDA Cleared Actigraph CENTREPOINT Insight Watch. Data was monitored during free living: sitting, walking, typing, drinking and during exercise: warming up, walking, running.

Results Study

Parameter

Corsano Algorithm (steps)

RMSE (Arms)

1.2

Bias  (+95% CI)

0.7

(0.6, 0.7)

95% LoA  (+ 95% CI) 

Lower

-2.5

(-2.5, -2.4)

95% LoA  (+ 95% CI)
Upper

3.8
(3.7, 3.8)

Steps Accuracy and Bias RadboudMC Study

Bland Altman Corsano Steps vs Actigraph

Note : the Activity (Steps) parameter is considered a software function intended to monitor and record daily energy expenditure and cardiovascular workout activities to allow awareness of one’s exercise activities to improve or maintain good cardiovascular health.

 

Conclusion / Claims

Given the bench and clinical testing, the following claims will be made:

Parameter

Range

Accuracy

Bias

PR

- Accuracy Arms <3 bpm

- Bias (+95%CI) ±0,5 bpm

Spo2

70-100%
 
70-80%
80-90%
90-100%

- Accuracy Arms <2 %
 
- Accuracy Arms <2 %
- Accuracy Arms <2 %
- Accuracy Arms <2 %

- Bias (+95%CI) ±0,5 %
 
- Bias (+95%CI) ±0,5 %
- Bias (+95%CI) ±0,5 %
- Bias (+95%CI) ±0,5 %

RR

Pooled
Supine
Prone
Lateral
Sitting 90°
Sitting 45°
Hyperventilation
Hypoventilation
Coughing
Treadmill

- Accuracy Arms <2 brpm
- Accuracy Arms <1 brpm
- Accuracy Arms <2 brpm
- Accuracy Arms <1 brpm
- Accuracy Arms <2 brpm
- Accuracy Arms <1 brpm
- Accuracy Arms <2 brpm
- Accuracy Arms <1 brpm
- Accuracy Arms <1 brpm
- Accuracy Arms <2 brpm

- Bias (+95%CI) ±0,5 brpm
- Bias (+95%CI) ±0,5 brpm
- Bias (+95%CI) ±1 brpm
- Bias (+95%CI) ±1 brpm
- Bias (+95%CI) ±1 brpm
- Bias (+95%CI) ±0,5 brpm
- Bias (+95%CI) ±0,5 brpm
- Bias (+95%CI) ±0,5 brpm
- Bias (+95%CI) ±0,5 brpm
- Bias (+95%CI) ±1 brpm

NIBP

- Accuracy Arms <5mmHg

- Bias (+95%CI) ±2 mmHg

Skin Temp

- Accuracy ±0.3°C (0,5 °F)

CBT

- Accuracy arms <0.5°C (0,9°F)

- Bias (+95%CI) ±0,2°C (±0,3°F)

Steps

- Accuracy Arms <2 steps

- Bias (+95%CI) ±0,8 steps