| T | Summary | Due Date | Status |
|---|---|---|---|
| REG_05: Casing material must be fire resistant | Done | ||
| REG_04: Marking on device and Label/Package | Done | ||
| REG_03: ESD protection | Done | ||
| REG_02: Battery protected against temperature | Done | ||
| REG_01: Battery is protected against over and under charge | Done | ||
| REQ_HW_37: Charging of the Bracelet shall not be physically possible during wearing of the bracelet (BLE active, HCP is both aware of charge level, and charging status on the webportal) | Done | ||
| REQ_HW_36: USB Charging cable must be dedicated custom designed magnetic connection | Done | ||
| REQ_HW_35: bracelet isolation shall be to BF as per IEC 60601-1 | Done | ||
| REQ_HW_34: AC Adapter shall be conform to IEC60601-1, dedicated & provided as part of the Corsano 287-2 System. | Done | ||
| REQ_HW_33: Packaging must be sufficiently resilent to protect the bracelet during multiple shipping | Done | ||
| REQ_HW_32: The IR LED power shall be less than 25mW/Sr at 20mA | Done | ||
| REQ_HW_31: The device shall use low drift accelerometer sensor | Done | ||
| REQ_HW_30: The device shall use low drift Temperature sensor | Done | ||
| REQ_HW_29: The External Devices shall be medically certified or approved (FDA cleared, MDR, etc.) | Done | ||
| REQ_HW_28: The Bracelet shall be compliant with EMC ISO 60601-2 requirements | Done | ||
| REQ_HW_27: The device is sealed to avoid easy access to the electronics | Done | ||
| REQ_HW_26: The device shall be able to operate as intended and safely at operating temperature | Done | ||
| REQ_HW_25: Electrode circuits shall comprise series resistors to limit the current | Done | ||
| REQ_HW_24: Case must be shock resistant | Done | ||
| REQ_HW_23: Charging status must be available | Done | ||
| REQ_HW_22: Charging cable must have a short-circuit protection | Done | ||
| REQ_HW_20: Device must have ESD protection, 15kV air, 8kV contact | Done | ||
| REQ_HW_19: Device must have non-volatile storage | Done | ||
| REQ_HW_18: Device must use AFE | Done | ||
| REQ_HW_17: Device must have an accelerometer | Done | ||
| REQ_HW_16: PPG interface must follow Philips reference design with G, R, IR | Done | ||
| REQ_HW_15: Battery must have max capacity | Done | ||
| REQ_HW_14: The device must have 3 leds for display | Done | ||
| REQ_HW_13: Casing material must be fire resistant | Done | ||
| REQ_HW_12: Marking on device and Label/Package | Done | ||
| REQ_HW_11: Battery protected against temperature | Done | ||
| REQ_HW_10: Battery is protected aginst over and under charge | Done | ||
| REQ_HW_09: Optical perfomance must be high to ensure good quality signals and medical grade measurements | Done | ||
| REQ_HW_07: The device should be easy to recharge and fool proof | Done | ||
| REQ_HW_06: The device must detect tap (by accelerometer) | Done | ||
| REQ_HW_05: The device must have a closed enclosure, water resistant | Done | ||
| REQ_HW_04: The system should be able to adjust precisely to the wrist to ensure good sensor position | Done | ||
| REQ_HW_03: The materials fo the device that are in contact with the skin should not create any allergy | Done | ||
| REQ_HW_02: The system should comfortable and be nice | Done | ||
| REQ_HW_01: The system should be based on low weight and small size | Done | ||
| USER_068: The app shall be able to run a third-party algorithm to monitor sleep | Done | ||
| USER_067: The app shall enable spot measurements of health metrics during a defined duration with manual entry option for HCP mode | Done | ||
| USER_066: The app shall collect raw sensor data (ACC, PPG, BIOZ, TEMP) and send it to the cloud with timestamps | Done | ||
| USER_065: The Corsano app in HCP mode shall generate and manage local technical alarms for connectivity and data update issues | Done | ||
| USER_064: The Corsano app in HCP mode shall display relevant health information and data update status | Done | ||
| USER_063: The Corsano app in Patient mode shall display relevant health information, connectivity status, and bracelet battery level | Done | ||
| USER_062: The mobile application shall handle metric and imperial unit systems for health data display | Done | ||
| USER_061: The mobile application shall run on and be compatible with supported operating system versions (iOS 14.5+, Android 8+) | Done | ||
| USER_060: The system shall be able to detect deviations from individual baseline values, allowing for a patient risk assessmentand early detection of clinical deterioration | Done | ||
| USER_059: The Corsano System shall be able to detect cardiac arrest based on the analysis of PPG signal | Done |